Principle and application of color measuring instrument based on color reproduction of Printing Books in China
Apr 27, 2023
Leave a message
Principle and application of color measuring instrument based on color reproduction of Printing Books in China
Color is the most critical element of printing and reproduction. How to ensure the true representation of color in the printing production process is a worldwide problem. At present, digital color management is adopted to describe the color characteristics of each production link in the production process, which is the best way to achieve accurate color reproduction, and its premise and foundation is the accuracy of color measurement data.
At present, density measurement has shown its superiority in some printing enterprises. Density measurement value can reflect the information of the ink layer thickness, and guide the control of dot expansion, through the density measurement data to control the printing quality, convenient and reliable, has been trusted by people. However, the density value can not directly reflect the stimulation of image color to the human eye. It can basically play a monitoring effect when it is used to monitor the printing color difference, but there are certain errors, the representation of color is not accurate enough, and it is not conducive to the communication of data with customers, so the measurement of excellence is derived. Chrominance measurement can better reflect the psychological and physiological laws of color vision, provide more detailed color information, and play a good role in promoting the printing color management.
Principle and application of densitometer
The densitometer consists of a light source, a lens group, a polarizer (optional), a color filter, a sensor and an electronic system, and a display.
Densitometer uses the built-in red, green and blue filters to measure the light reflection or transmission of yellow, magenta and green colors, and calculates the density value. This principle based on the three-color filter makes its structure very simple and widely used, but because of the defects of the filter itself, it also constitutes the insuperable limitations of the densitometer: It can only measure the density value of printed matter, but does not include the expression of hue, so it can not truly reflect the actual visual effect of color, which is a kind of "color blindness" measuring equipment.
In addition, densitometer measurement also has some other limitations, such as: the field of application is limited to the four-color printing process, although often used to assist in monitoring ink thickness, but there is no direct correlation between density and ink thickness. Therefore, the densitometer is used for users to provide the maximum/small density, dot expansion and printing contrast on the film or plate for pre-press compensation and correction, to guide the production management personnel to correctly screen, determine the amount of ink, exposure, ink balance and other control parameters, and is not good at color management color reproduction accuracy measurement and control.
Two, photoelectric colorimeter principle and application
The photoelectric colorimeter can be regarded as a reflectance meter with a special set of three filters, which are different from the red, green and blue filters of the densitometer. This filter weights the wavelengths of the spectrum in each channel of the colorimeter according to the CIE spectral tri stimulus value. It is primarily a reflectance problem, not a logarithmic problem.
Compared with density measurement, photoelectric colorimeter can specifically describe color information through three stimulus values, rather than only limited to brightness information. Because it still adopts the principle of three color filter, the spectrum range of sampling is limited, so the accuracy is not high, which is not suitable for color measurement and control in high-precision color management.
Principle and application of spectrophotometer
Spectrophotometry is to measure the spectral reflection of the whole visible spectrum at equal intervals. Compared with photoelectric colorimeter, spectrophotometry can be regarded as continuous spectrum measurement, it provides much more color information, much richer.
There are three basic spectrophotometer spectrophotometry principles: rotating color filter spectrophotometry, scattering prism spectrophotometry and diffraction grating spectrophotometry, as shown in Figure 3. The first method is to install 20 to 30 narrow-band color filters on the disc, by rotating the disc to achieve light separation. The latter two methods use the dispersion of light to decompose the composite radiation of the light source into monochromatic radiation of different wavelengths, and arrange it in a certain order. The dispersive elements used are prisms or diffraction gratings.
For example, the scanning spectrophotometer on CPC2 in Heidelberg is based on the principle of diffraction grating.
Common spectrophotometers have wavelength intervals of 10 or 20nm, and the visible spectrum being recorded is divided into about 30 segments. In some high-precision systems, the measurement interval can also be smaller (up to 1nm). Then, under the specified illumination body and observation field, the light quantity of each wave length is measured one by one through the photodetector, and then the chrominance value of the product to be measured is calculated according to the reflection spectrum or transmission spectrum.
The CIE coordinates of different illuminated objects and observed fields can also be calculated from the spectral measurements of a spectrophotometer. Specifically, when converting from spectral values to CIE color trististimulus values, the illuminating body is directly taken as a parameter, so mathematical approximation is used when converting from one illuminating condition to another.
A spectrophotometer is closer to the visual response of the human eye than either a densitometer or a colorimeter, because it measures the amount of light reflected across the visible spectrum, but it is different from the human eye. The eye evaluates light on the basis of feeling all wavelengths at the same time, while the reflection spectrum of printed matter must be measured wavelength by wavelength, which means the spectrum must be decomposed at each wavelength before the spectrum enters the photoelectric receiver. At present, most spectrophotometers use the reflected light path for spectrophotometry, and it is not a real spectrophotometer. It's just a sum, not an integral, for a predetermined wavelength measurement.
The color defined by spectral data is more complete, the measurement accuracy is high, and spot color can be measured, and the spectral data can be calculated to obtain the density value and chroma value. It is suitable for spot color evaluation, spectral analysis and color evaluation in the process of color management, as well as the production of equipment color characteristics file.
Theoretically, all inks, whether four-color ink or spot color ink, can be measured by spectrophotometer. The system automatically compares the measured data with the target color value, and displays the result on the screen. If the density value is selected, the mass can be determined by traditional methods; If the Lab value is selected, the color deviation can be visually determined by the △E value, and the amount of deviation will be shown in a chart. The operator can use the data on the chart to determine which area is the correct color, and which area has more or less ink. If the operator decides to make a color adjustment, the spectrophotometer is also controlled online, with the press of a button sending the recommended adjustment data to the ink zone Settings.
If the density value can not express a spot color, the same is true for four color ink. This is because the density measurement and ink pigment content has no relationship, the density value does not contain the expression of hue, which is a problem that all printing enterprises are familiar with. If the type of ink changes, although there is the same printing density, it is not necessarily guaranteed to obtain the same printing color.
The spectrophotometer also functions as a densitometer, and density measurement is not a technique that can be completely abandoned overnight. Although some printers have moved to the new technology, they still use the old technology as a tool to judge quality, which is well suited to the diverse needs of users.
In addition to controlling the amount of ink in the ink area during the printing process, spectrophotometry has several other advantages. In the process of color management, in order to ensure the accuracy and consistency of the color in each link of the printed matter from input to output, we need to use the ICC Profile Profile profile for color management in each link. The spectrophotometric measurement method provides the premise and guarantee for the production of profile profile profile: Because Profile only supports Lab or XYZ values, not density values.
Iv. Actual use
Currently, the most popular instrument in the market is X-Rite's 500 Series spectrophotometer, which covers density measurement, dot increase, overprint rate, print contrast, gray balance, tone error, color comparison and chrominance functions. The 530 can also be used with color quality control software; X-Rite's DTP41 automatic scanning spectrophotometer is also used by many users and has a high degree of automation, making it easy and fast to create ICCProfile feature description files.
GretagMacbeth's SpectroEye portable spectrophotometer provides all of the chrominance measurement capabilities needed to accurately measure and control spot colors, as well as all of the chrominance density measurements used to monitor colors during printing. A wide range of color Spaces and color difference formulas are provided by SpectroEye, including CMC, FMCⅡ, or ΔE*94 color difference formulas for CIELAB color systems to ensure perfect compliance with standard procedures for instrumented color quality control, and can even be used as part of the ink color matching system for ink matching.
In addition, EFI's ES-1000 spectrophotometer is easy to measure both monitor screens and printed matter. Normal color management software can support it, and users can choose to measure by a single color block or a single color bar as required. Some models can also support off-line battery-powered mode, this mode EFI spectrophotometer can measure up to 512 color block readings, convenient to upload to the computer, suitable for off-line operation.
V. Concluding remarks
In general, now the colorimeter is used relatively less, densitometer and spectrophotometer has shown their own advantages, although the former has no chroma function, but reflect the ink layer thickness, monitoring the printing state, throughout the printing process can be used, and the price is relatively cheap; Spectrophotometer has both density function and chroma function, and high precision, is the future color reproduction color management trend, but at the same time the price is high, has not been widely popular.
However, no matter which kind of measuring equipment is used, it is very important to correct the equipment calibration and parameter setting, only in this way can ensure the normal use of the equipment and the reference value of the measured value. Generally, the measurement parameters should be considered include standard light source, measurement aperture, field Angle, response mode, calculation formula and coefficient, etc. For density measurement, the selection of three-color filter and polarized filter should be considered. In the color management process, it is best to use the same measuring instrument and the same color management software. For example, the same ES-1000 spectrophotometer is used in the digital proofing linearization and ICC Profile production processes to ensure that the color descriptions are comparable and thus reduce the error.