Table Game Of Cards Printing color measurement should pay attention to some problems
Mar 21, 2023
Leave a message
Table Game Of Cards Printing color measurement should pay attention to some problems
Density measurement has been the most commonly used measurement method in the printing industry, but densitometer can not provide psychophysical quantities related to the sensitivity of the human eye, so its analytical measurement ability is limited. Therefore, in the detection and evaluation of printed matter, chrominance measurement plays an increasingly important role. In color measurement, the following problems must be paid attention to in order to get the measurement data that can correctly reflect the quality of printed matter.
1. Correct the whiteboard
Since densitometers and spectrophotometers are very precise instruments, routine calibration of measuring instruments should be carried out before measurement, so as to ensure the accuracy of measurement results. Typically, each unit comes with a standard whiteboard, such as the DensiEye750 densitometer, X-R ite500 series spectrophotometer, etc. According to the purpose of data measurement, we choose to zero on the white board or on the white paper, set the correction warning, if the correction time is exceeded, the white board should be corrected.
2. Substrate
The color measurement value will be different due to the different substrate used in the measurement. The effect of the two kinds of substrate on the color measurement will be different due to the different transparency of the substrate, and the greater the transparency of the substrate material affected by the greater. Therefore, in the process of high quality printing production and its color matching inspection, attention should be paid to the placement and use of the correct substrate on the test table.
The following criteria are generally recommended: when the substrate opacity is greater than or equal to 99, the measurement results will not be affected by the substrate, when the opacity is between 95 and 99, the color measurement should use black substrate, and when the opacity is less than 95, use white substrate.
In actual production, people usually use white substrate as the substrate for color measurement. For example, when measuring and inspecting the color of packaging printed matter such as plastic film with greater transparency, white substrate in line with ISO standard should be placed under the tested printed matter to avoid unnecessary errors. At the same time, when measuring the color information on the same substrate material, we should pay attention to the selection of the same substrate, that is, we should pay attention to the consistency of substrate use.
3. Measure the light source
A, B and C are simulated incandescent lamps, midday daylight, cloudy daylight or cloudy noon daylight, respectively, while the radiation distribution of D65 light source is obtained through a complex averaging process after many measurements of the daylight spectrum at different times, climates and locations. A light source and other light source measurement results are very different, now rarely used, F series light source is generally used to measure fluorescent products, we can say that C light source and D65 light source is the most useful for the printing industry. D series light source because its color temperature is 5000K-7500K, close to white, so its display is good, in the printing industry, it is recommended to use D50 light source when observing transmission samples, and D65 light source when observing reflection samples. Therefore, the data must be measured under what light source must be indicated.
4. Measure the Angle of view
The color resolution of human eye is affected by the size of the field of view. Experimental results show that the ability of human eye to distinguish color difference is low when viewing color with a small field of view (< 4°). When the field of view increases from 2° to 10°, the accuracy of color matching and the ability to distinguish color difference are increased. But when the field of view is further increased, the accuracy of color matching is not much improved. Under 2° field of view and 10° field of view, the same color will present different colors. Therefore, in color measurement, the field Angle selected for measurement must also be marked.
5. Color space
Standard colorimetry system has a variety of color space, that is, a variety of different expression forms can be selected, the choice of color space is actually to choose the expression form of measurement results.
6. Measurement of color difference
Color difference is an important parameter index of print quality detection and evaluation. The size of color difference directly affects the quality and grade of products. The color difference formula based on CIE1976LAB uniform color space is commonly used in the printing industry, but the data expression of the color difference formula cannot match the visual perception of human eyes. Therefore, color experts have successively launched CMC (l: c) Color difference formula, CIE94 color difference formula and the latest CIEDE2000 color difference formula, among which CMC (l: c) color difference formula is listed as the international and national standard in the textile industry.
Measurements using different color difference formulas vary widely. Some people measured the tri stimulus value of 10 pairs of color samples by experiment, and found that there was a trend of △ELAB > △ECMC(2:1) > △E2000(1:1:1) by comparing the color difference calculated by three different color difference formulas. Therefore, in the measurement of color difference data must also be noted which color difference formula. [next]
7. Aperture size
When it comes to color measurement, the aperture of the color measuring instrument may be larger than the edge of the measuring color block. At this time, it is necessary to choose other color blocks for measurement or other color measuring instruments. The aperture size specified by the user should not be larger than 5mm, generally with standard 3 ~ 8mm and small 1 ~ 7mm aperture. The aperture of 3 ~ 8mm is used to measure the color scale used in standard sheet printing (control strip), and the aperture of 1 ~ 7mm is used to measure the color scale used in web printing.
8. Response mode
Response states include T, E, A, I and so on. State T and state E are the two most commonly used states. The commonly used wide band noise should be T-status (American ANSI standard, widely used in the North American printing industry), and the commonly used narrow band noise should be E-status (European DIN standard). When E status is selected, the densitometer is more sensitive to small changes in the print density value. In the aspect of print measurement, when state T or state E is used, for yellow, the value measured under the state E is larger than that measured under the state T. Therefore, it is necessary to adjust the response state to be consistent when comparing and evaluating the measured values.
9. Observation conditions
The 0/d condition means that the light source basically illuminates the sample vertically, and the diffuse light of the sample is received by the integrating sphere. In the case of regular reflection, the reflected energy of the sample is fully received, which is the "reflection ratio" in the true physical sense. d/0 condition refers to that the light diffused by the integrating sphere illuminates the sample, and receives reflected light basically perpendicular to the surface of the sample. This condition is closer to the observation of the object by human eyes in general, that is, the sample is illuminated by white light from all sides, and human eyes are basically perpendicular to the sample for observation. The chrominance values of samples are different under different lighting/observation conditions.
The sample with good diffuse reflection is little affected by geometric conditions, while the sample with poor diffuse reflection has high glossiness. The difference of geometric conditions will bring about the difference of light flux received by the detector, and its reflection factor is greatly affected by geometric conditions. Samples measured under different geometric conditions will not change the proportion of the tri stimulus values, that is, different geometric conditions have almost no effect on chromaticity coordinates, but the spectral reflection factor is different, resulting in different tri stimulus values. Therefore, as long as it is not an ideal Lambert body, the difference in geometric conditions will affect the spectral reflection ratio and total reflection ratio of the sample under test, and the degree of influence is related to the luster of the sample surface.
The surface of white paper has good diffusivity, weak luster and high reflection factor. Inferred from the above, its chroma value should be close to that of different geometric conditions. The data suggest that is not the case. The reason for this is that the white paper has been added with fluorescent brightening materials, so its test has a special feature.
Fluorescent materials have fluorescence excitation characteristics. When a certain or a certain wavelength of light is used to irradiate fluorescent materials, the fluorescent materials will be excited with light longer than the irradiation wavelength. The outgoing light contains both the reflected part of the irradiation light and the fluorescent radiation part excited by the irradiation light. Therefore, the position of the monochromator in the measuring device will have a great influence on the measurement results. When measuring the chrominance value of a fluorescent sample, a post-spectrophotometer should be used, that is, the monochromator is located behind the sample and before the detector.
The non-fluorescent sample has no fluorescence excitation characteristics, and the light incident on it follows the law of reflection, with only reflected flux but no radiative flux. No matter whether the monochromatic light is illuminated or received, what the detector receives is the reflected flux of the sample to the irradiation wavelength, and the position of the monochromator has no influence on the measurement.
In principle, color measuring instruments with different illumination/observation conditions cannot be substituted for each other, especially for high-gloss samples and fluorescent materials. Therefore, when conducting chrominance tests, it should be clear what lighting/viewing conditions to adopt. When purchasing a color measuring instrument, check whether the lighting/viewing conditions of the instrument are consistent with the corresponding product standards.
10. Summary
In the process of chrominance measurement, we must pay attention to the selection of substrate, light source, color space and so on, and should be noted in the measurement results, otherwise the measurement results are likely to be inconsistent with the customer's parameter index requirements, so that should be qualified products into unqualified products, is not conducive to the enterprise production process data, standardized implementation.